- Off: A single baseline is drawn from the onset of the first peak of
the fused group to the termination of the last peak of the group.
(Default)
- On: Causes a separate baseline to be drawn for each peak in the
fused group.
Negative Peak: Determines whether peak detection will detect
inverted peaks, which are peaks that point downward from the
baseline. At any given moment we can detect positive or negative
peaks but not both at once.
- Off: Detect positive peaks. (Default)
- On: Detective negative peaks.
TEV
Type
SW Auto Zero: Re-zeros the baseline of the trace at the specified time for
the specified detector. Used after a FID gain change event or a spectrum
gain change event.
Note
The Single Base and Fused Override events can act together to produce
multiple horizontal baselines, at different heights, for a fused peak group.
Value The values available depend on the integration type selected from the
TEV Type column.
Slope Sensitivity and Peak Width: Enter the number of points,
between 1 and 99, to be used.
Single Baseline: Select Off, End, Bgn.
SW Auto Zero: No options.
All other integration types: Select On or Off.
Det # The ID number of the detector that will be affected by the event. Valid
values are 1 and 2.
Time Indicates the time, in seconds, that the event should occur during the
analysis. Enter a value between 0.0 and 3600.0.
Note
Event times must be less than the analysis time.
4. To save the changes and close the window, click OK.
Note
To save the changes without closing the window, click Save.
Application
84