Calibration
Repeatability
Check
If enabled, the GC will perform a check of the repeatability of
calibration runs to the limits specified in ISO6974-1984(E) 8.2.1
Table 6. If the calibration fails to meet the conditions set forth in
the table, then the calibration is deemed to have failed and the GC
will rerun the calibration sequence.
Metrology Type Shows the metrology type that the GC is configured for.
GC Id
Identification
Number
Configuration
Checksum at
Lockout
The checksum of the configuration fields that is calculated when
the security switch is locked.
Current
Configuration
Checksum
The GC will periodically recalculate and update the configuration
checksum. This ‘current’ value will be the latest calculated value.
Checksum Update
Time
The time that the configuration checksum was last updated.
GC Mode Allows you to select an operating mode for the GC. See Section 1.8 for
more information.
Default Stream
Sequence
Sets the default sequence to be used by the indicated detector during
auto-sequencing. To create a new stream sequence or to edit an
already-created sequence, click Stream Sequence....
See Section 4.10 for more information.
After making changes, click Save to save the changes without closing the window. To save
the changes and close the window, click OK.
4.2 The Component Data Tables
MON2020 allows you to view and edit the component data tables. The number of available
component data tables depends on the GC unit configuration.
To assign a component data table to a stream, see Section 4.9.2.
1. To view a component data table, select Component Data... from the Application
menu.
The Component Data Tables window appears, displaying a list of available
component data tables.
Note
Other ways of accessing the component data tables are by pressing F6 or by clicking
from the Toolbar.
2. Select the table that you want to view.