Emerson MON2020 Gas Grill User Manual


 
Print at end of
run
Prints the chromatogram to the PC's default printer at the end of the
run and is unchecked by default.
Save at end of
run
Saves the chromatogram to the GC's Data folder at the end of the run
and is unchecked by default.
Show bunched
data
If this box is unchecked, then all of the raw data points are plotted to
the chromatogram window; if this box is checked, which is the default
option, then each point plotted on the graph represents the average
of a group of raw data values. The size of the data group is determined
by the peak width value listed in the Timed Events table.
2.5.1 The Chromatogram Viewer's Timed Events table
The Chromatogram Viewer displays a compact version of the Timed Events table, located
on the upper right side of the window. The events displayed in the table are sorted by time.
See Section 4.3 for more information.
The Timed Event table displays the following data for each event:
Event Type The type of timed event. These events are mapped to the Time Events
window and include Valve, Integration and Gain events.
Vlv/Det Identifies which valve or detector is involved in the event.
Value Setting of the event; for example, a valve was turned ON, or the gain was
set to 4.
Time (s) The number of seconds into the cycle that the event occurred or will occur.
Timed events from live or archived chromatograms can be edited from the Chromatogram
Viewer by double-clicking on the Timed Events table. The changes will affect the next
analysis run. The following commands are available by right-clicking on the table:
Auto Scroll When checked, if a live trace has been selected from the
Chromatogram bar's pull-down menu, the Timed Event table will keep
its focus on the event closest in time by highlighting that event in dark
blue.
Save Sheet Allows you to save the table to the PC in one of the following formats:
TXT, CSV, XLS, HTM, or XML.
Copy to
Clipboard
Allows you to copy the table to the clipboard.
This data can be pasted into another application such as Microsoft
Word or Microsoft Excel.
Chromatograph