Emerson MON2020 Gas Grill User Manual


 
Note
If auto calibration or auto validation is enabled then they will be performed as part of the auto
sequence.
To start an analysis run on a single calibration or sample stream, do the following:
1. Select Single Stream… from the Control menu.
The Start Single Stream Analysis dialog displays.
2. Select a stream from the Stream menu.
3. Decide whether to enable purging, then select or unselect the Purge stream for 60
seconds check box as necessary.
Purging allows sample gas to flow through the sample loop for 60 seconds prior to
beginning the first analysis. The checkbox is selected by default.
4. Check the Continuous operation check box to set or disable repetitive analysis. The
checkbox is selected by default.
5. Click OK.
The analysis starts. Use the Mode column on the GC Status Bar to monitor the status
of the analysis run.
6.3 Calibrate the gas chromatograph
Calibration runs are determined by the CDT and Streams settings. See Section 4.2 and
Section 4.10 for detailed instructions on how to edit these settings.
To calibrate a GC, do the following:
1. Select Calibration… from the Control menu.
The Start Calibration dialog displays.
Note
If the GC is in Auto Sequence mode, calibration will not start until two or more analysis runs
have been completed. This delay is required to complete the current analysis and the analysis
of the stream currently purging through the valve.
2. Select a stream from the Stream menu.
3. Decide whether to enable purging, then select or unselect the Purge stream for 60
seconds check box as necessary.
Purging allows sample gas to flow through the sample loop for 60 seconds prior to
beginning the first analysis. The checkbox is selected by default.
4. Select the desired calibration type.
Analysis
148