6.2.3 Setting of calibration range
This mode is used to set if the range of each CH
(component) at the calibration manual calibration or auto
calibration) should be calibrated with a single range or 2
ranges.
1. During measurement, press the MODE key to display
the User mode.
2. Point the cursor to “Setting about Calibration” by
pressing the UP or DOWN key. Press the ENT key.
3. In the “Setting about Calibration” screen that appears,
point the cursor to “About Calibration Range” by pressing
the UP or DOWN key. Press the ENT key.
4. In the “Calibration Range CH Selection” screen that
appears, point the cursor to the CH you want to set by
pressing the UP or DOWN key. Press the ENT key.