Emerson 8712E Grill Accessory User Manual


 
Quick Installation Guide
00825-0100-4664, Rev BB
January 2013
Rosemount 8712 / 8700 Series
8
The electrodes in the Rosemount 8705 sensor are properly orientated when the two
measurement electrodes are in the 3 and 9 o’clock positions, as shown on the right of
Figure 4.
The electrodes in the Rosemount 8711 are properly orientated when the top of the sensor is
either vertical or horizontal, as shown in Figure 5. Avoid any mounting orientation that
positions the top of the sensor at 45° from the vertical or horizontal position.
Figure 4. Sensor Orientation
Figure 5. Rosemount 8711 Mounting Position
FLOW
FLOW
45° Electrode Plane
45° Electrode Plane
4664RevBBQIG.fm Page 8 Friday, January 11, 2013 6:13 PM